Wear debris from ultra-high-molecular-weight polyethylene components used for joint replacement prostheses can cause significant clinical complications, and it is essential to be able to predict implant wear accurately in vitro to prevent unsafe implant designs continuing to clinical trials. The established method to predict wear is simulator testing, but the significant equipment costs, experimental time and equipment availability can be prohibitive. It is possible to predict implant wear using finite element methods, though those reported in the literature simplify the material behaviour of polyethylene and typically use linear or elastoplastic material models. Such models cannot represent the creep or viscoelastic material behaviour and may introduce significant error. However, the magnitude of this error and the importance of this simplification have never been determined. This study compares the volume of predicted wear from a standard elastoplastic model, to a fractional viscoelastic material model. Both models have been fitted to the experimental data. Standard tensile tests in accordance with ISO 527-3 and tensile creep recovery tests were performed to experimentally characterise both (a) the elastoplastic parameters and (b) creep and relaxation behaviour of the ultra-high molecular weight polyethylene. Digital image correlation technique was used in order to measure the strain field. The predicted wear with the two material models was compared for a finite element model of a mobile-bearing unicompartmental knee replacement, and wear predictions were made using Archard’s law. The fractional viscoelastic material model predicted almost ten times as much wear compared to the elastoplastic material representation. This work quantifies, for the first time, the error introduced by use of a simplified material model in polyethylene wear predictions, and shows the importance of representing the viscoelastic behaviour of polyethylene for wear predictions.

Viscoelastic material models for more accurate polyethylene wear estimation / Alotta, G.; Olga, Barrera; Pegg, E. C.. - In: JOURNAL OF STRAIN ANALYSIS FOR ENGINEERING DESIGN. - ISSN 0309-3247. - 53:5(2018), pp. 302-312. [10.1177/0309324718765512]

Viscoelastic material models for more accurate polyethylene wear estimation

G. Alotta;
2018-01-01

Abstract

Wear debris from ultra-high-molecular-weight polyethylene components used for joint replacement prostheses can cause significant clinical complications, and it is essential to be able to predict implant wear accurately in vitro to prevent unsafe implant designs continuing to clinical trials. The established method to predict wear is simulator testing, but the significant equipment costs, experimental time and equipment availability can be prohibitive. It is possible to predict implant wear using finite element methods, though those reported in the literature simplify the material behaviour of polyethylene and typically use linear or elastoplastic material models. Such models cannot represent the creep or viscoelastic material behaviour and may introduce significant error. However, the magnitude of this error and the importance of this simplification have never been determined. This study compares the volume of predicted wear from a standard elastoplastic model, to a fractional viscoelastic material model. Both models have been fitted to the experimental data. Standard tensile tests in accordance with ISO 527-3 and tensile creep recovery tests were performed to experimentally characterise both (a) the elastoplastic parameters and (b) creep and relaxation behaviour of the ultra-high molecular weight polyethylene. Digital image correlation technique was used in order to measure the strain field. The predicted wear with the two material models was compared for a finite element model of a mobile-bearing unicompartmental knee replacement, and wear predictions were made using Archard’s law. The fractional viscoelastic material model predicted almost ten times as much wear compared to the elastoplastic material representation. This work quantifies, for the first time, the error introduced by use of a simplified material model in polyethylene wear predictions, and shows the importance of representing the viscoelastic behaviour of polyethylene for wear predictions.
2018
finite element analysis; fractional viscoelasticity; material model; Polyethylene wear; unicompartmental knee arthroplasty; Modeling and Simulation; Mechanics of Materials; Mechanical Engineering; Applied Mathematics
File in questo prodotto:
File Dimensione Formato  
Alotta Barrera Pegg 2018 preprint.pdf

accesso aperto

Tipologia: Documento in Pre-print
Licenza: Dominio pubblico
Dimensione 3.28 MB
Formato Adobe PDF
3.28 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12318/47175
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 12
social impact